10 MBB-M-BA-e-1314 5 Appliance overview 213 4 5 1. Weighing pan 2. Bubble level 3. Display unit 4. Stand (optional) 5. Wall bracket
MBB-M-BA-e-1314 11 6 Overview of displays Display Designation Description 1 0 Zeroing display Should the balance not display exactly zero
12 MBB-M-BA-e-1314 7 Keyboard overview Button Designation Function ON/OFF-switch Turn on/off Zero setting key Balance will be reset to 0.0
MBB-M-BA-e-1314 13 8 Transportation & Storage 8.1 Testing upon acceptance When receiving the appliance, please check packaging immediately
14 MBB-M-BA-e-1314 9 Unpacking, Setup and Commissioning 9.1 Installation Site, Location of Use The balances are designed in a way that rel
MBB-M-BA-e-1314 15 9.3 Scope of supply Balance Power unit (in conformity with EN 60601-1), including fuse and LED Wall bracket Oper
16 MBB-M-BA-e-1314 3. Then screw in the corresponding locking screw to achieve a tight fit. 4. Use the supplied knurled screws to attach t
MBB-M-BA-e-1314 17 9.6 Stand assembly (optional) Use the 4 screws to attach the stand to the bottom of the scale. Remove the two latera
18 MBB-M-BA-e-1314 9.8 Rechargeable battery operation (optional) When an optional rechargeable battery is used, proceed as follows: On model
MBB-M-BA-e-1314 19 Carefully insert the rechargeable battery block and insert plug into connection CN 3 Ensure that the cables are not squee
2 MBB-M-BA-e-1314 GB KERN MBB-M Version 1.4 02/2013 Operating instructions Baby weighing scale Table of Contents 1 Technical Data ...
20 MBB-M-BA-e-1314 9.9 Connecting the power supply Power is supplied by the external power unit which also serves to isolate the mains supply
MBB-M-BA-e-1314 21 10 Menu overview Start balance by pressing OFFON Press for 3 sec., “SETUP“ will be displayed. Select paramete
22 MBB-M-BA-e-1314 11 Operation 11.1 Weighing 0 Start balance by pressing OFFON The balance carry out a segment test, then the program
MBB-M-BA-e-1314 23 The balance is able to only store one taring value. When the balance is unloaded the saved taring value is displayed w
24 MBB-M-BA-e-1314 12 Error messages Display Description Underload Weight on weighing pan is too low. Please increase weight. If the erro
MBB-M-BA-e-1314 25 13 Service, maintenance, disposal 13.1 Cleaning Disconnect the unit from the mains power supply prior to cleaning. Please
26 MBB-M-BA-e-1314 14 Instant help In case of a fault in the program sequence, the balance should be shortly switched off. The weighing proces
MBB-M-BA-e-1314 27 15 Verification Verified scales bear a verification mark or one or more seals affixed by the Bureau of Standards or the man
28 MBB-M-BA-e-1314 Press HOLD repeatedly until “CAL 0“ appears. Press TARE, the display will show a numeric value. Then
MBB-M-BA-e-1314 29 15.2 Adjustment controls and seals After a verification the balance is sealed at the indicated positions. Verification of th
MBB-M-BA-e-1314 3 9 Unpacking, Setup and Commissioning ... 14 9.1 Installation Site, Locatio
30 MBB-M-BA-e-1314 Position of the adjustment switch: Position of the adjustment switch State 1. to left Not documented 2. concentric
MBB-M-BA-e-1314 31 15.4 Navigation in the menu With the balance switched on, keep the button depressed for approx. 3 sec until "SET
32 MBB-M-BA-e-1314 Press the [HOLD] and [TARE] buttons to make the selection * default setting * * * * * *
MBB-M-BA-e-1314 33 Description: Weighing unit: kg Partition steps, weighing range (max.) and readout (d) Multi-range balance/ single-range ba
34 MBB-M-BA-e-1314 15.5 Verification validity period (current status in G) Personal scales in hospitals 4 year Personal scales if not
4 MBB-M-BA-e-1314 1 Technical Data KERN MBB 15K5M MBB 15K2DM Display 6-digit Weighing range (max) 15 kg 6 kg; 15 kg Minimum load (Min) 100 g
MBB-M-BA-e-1314 5 2 Declaration of conformity Declaration of conformity: See separate document showing serial number of device CE marking:
6 MBB-M-BA-e-1314 “Electro-medical appliance“ with attachment for type B Device protection category II Dispose of old appliances
MBB-M-BA-e-1314 7 3 Basic Information (General) Weighing instruments have to be verified for the purposes stated below in accordance with
8 MBB-M-BA-e-1314 3.3 Improper Use Do not use these scales for dynamic weighing processes. Do not leave permanent load on the balance
MBB-M-BA-e-1314 9 4 Basic Safety Precautions 4.1 Pay attention to the instructions in the Operation Manual Carefully read this operation manu