64 MLS_C-BA-e-1330
How to enter the determined corrective factor:
Press the arrow keys ( ), to select “Corrective
Factor“
Press the -button and the current setting will flash
Press the arrow keys, to enter the determined corrective
factor
Use the , to select the digit
Press the keys, to select the digit on the left/right
(relevant enabled digit flashes).
Confirm setting by pressing the PRINT-key and enter
additional parameters as follows.
9. Standard setting
This setting is used to set all drying parameters to factory settings.
Press the arrow keys ( ), to select “Standard“
Press the -key; Query “Standard?“ appears
Confirm by pressing the PRINT-key; drying parameters
are reset to factory settings.
The device returns to the menu.
If the drying parameters are not reset to factory settings,
press the ESC-key.