12.5 Service menu
Configure display unit:
Single-range balance:
Weighing scale is turned off
Press and hold;
Keep holding pressed down and also hold down
at the same time until software version “P1.0x” is
displayed. Release both keys.
“SCL.TYP”will appear briefly followed by “1.rAnGE“.
Setting single range (“1.rAnGE“) – or dual range
(“2.rAnGE“) weighing scale
Press to change between “1.rAnGE”and “2.rAnGE”.
Select „1.rAnGE“ and confirm by .
„bASUnt“ is displayed. This is where you can set the required
weighing unit.
Press to change between the individual units and
press to select the required unit (ex. here: „kg“).
“SCL.CAP” will appear briefly.
Setting weighing scale capacity
A numeric value appears and the digit on the right is
flashing.
Press to increase the numeric value or to
decrease. Press to change to a different decimal
place. Set weighing scale capacity (here is an example of
50 kg) and press to confirm.